The Thermo Scientific TriPlus 300 Headspace autosampler offers a comprehensive, flexible solution compatible with standard gas chromatographs (GC) and GC/MS systems, including those using packed and capillary columns, accelerating your organic volatiles determination workflow while minimizing down-time. The autosampler enables users to analyze a larger number of samples more quickly and achieve unparalleled productivity with enhanced sample tray and incubation oven capacity.
Manufacture Specifications
Compatible carrier gas:
- Hydrogen, helium, nitrogen
Auxiliary gas:
- User-selectable helium and nitrogen
Auxiliary gas electronic flow control:
Flow rate: 10 to 300 mL/min
Precision: /- 0.1 mL/min
Safety control: On auxiliary gas presence
Maximum inlet pressure: 9 bar
Vial pressure: Up to 4 bar
Pressure settings:
- bar, psi, kPa
Carrier gas electronic control option:
Pressure range: 0–120 psi
Carrier gas flow rate: max 1000 mL/min (He)
Compatible gas types: He, N2H2 , Ar, Ar 5%CH
Compatible columns: Packed columns, wide-bore and capillary columns
Operation modes: TriPlus 300 HS control (direct control),
TriPlus 300 HS GC mode (additional flow)
Pressure/flow programming: Constant or programmed flow, constant or programmed pressure.
Flow programming: 24 ramps
Pressure programming: 24 ramps
Pressure setpoint increment: 0.01 psi
Flow setpoint increment: 0.1 mL/min
Atmospheric pressure and
temperature compensation: Yes
Synchronization with GC oven: Up to 25 oven ramps and 26 isotherms