Used for quantitative and qualitative analysis of the elemental composition of metal alloys, liquids, food and biological samples, minerals etc...
Element detection rangefrom sulfur to Uranium
Data acquisition time10 1200 s
Power supply 90 - 240 V, 50/60 Hz
Power consumption40 W
Dimensions430 mm x 340 mm x 200 mm
Weight18 kg
X-ray tubeW or Ag anode
250 ?m Be window
air-cooling
X-ray generatorvoltage: 450 kV, adjustable in 100 V steps
current: 0100 ?A, adjustable in 0.2 ?A steps
maximal power: 5 W
Stability0.1% pr 8 hours
Detector type SDD with thermo-electrical cooling
Active area 6 mm2
Energy resolution 125 eV
Pulse processordigital pulse processor
pile-up rejection
pulse shape selector
automatic adjustment to count rate
ADC4096 channels
Buffer RAM4096 32-bit channels
Timersreal and "live" time
Supported operating systemsWindows 7 or 10
Quantitative analysis algorithmsfundamental parameters
quadratic stepwise multiple regression
manual spectra comparison