Dimensions:
- Size: (H x W x D) 1060 x 465 x 525 mm
- Weight Approx.: 95 kg empty, 121 kg full
- Noise level (normal operation): 38 Lpa dB
Temperature Control:
- Control range: 1C to 40,5C
- Setting resolution: 0.1C
- Temperature measurement accuracy: 0.3C
- Temperatures display range: 0C to 99C
Cooling system:
- Tank volume: 26 litres
- Initial cooling capacity: (relative 20C) 6250 kJ
- Continuous cooling capacity: 4270 kJ/h (1186 W)
Circulation system:
- Flow capacity, main circuit: Approx. 21.0 l/min at zero head (adjust-able)
- Flow capacity, cardioplegia circuit: Approx. 7.0 l/min at zero head
- Maximum pressure, main circuit: Approx. 2.4 bar (adjustable)
- Maximum pressure, cardioplegia circuit: Approx. 0.9 bar