Description
- Model: BK-6310
- :
- : Throughput
- 60Tests/hour:
- :
- Assay Items: 5 parts, 29parameters,3 histograms
- :
- : Principle
- Tri-angle laser scatter, Flow cytometrymethod, 3D scattergram analysis, Impedancemethodfor RBC and PLT counting, Cyanide-free method for HGB test:
- :
- TestMode: CBC mode,CBC+DIFF mode Venous whole blood, Capillary whole blood and Prediluted.
- :
- : Parameters
- WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PCT,PDW, P-LCR, P-LCC, NEU%, LYM%, MON%, EOS%,BAS%,NEU#, LYM#, MON#, EOS#, BAS# 4 Research parameter: ALY%, ALY#, IG%, IG#:
- :
- Performance: Item
- Linearity range: Carry Over
- CV:
- WBC: (0~300.0x10^9)/L
- ≤0.5%: ≤2.0%
- : RBC
- (0~8.00x10^12)/L: ≤0.5%
- ≤1.5%:
- HGB: (0~250)g/L
- ≤0.5%: ≤1.5%
- : PLT
- (0~3000x10^9)/L: ≤1.0%
- ≤4.0%: Sample Volume
- CBC+DIFF mode≤20μl CBC mode≤10μl:
- :
- Storage: 10,000 resultsincludinghistograms,scattergrams and patient information
- :
- : Interface
- 4 USB ports, 1LAN port Bi-direction LIS, support HL7 protocol,Internal RFID reader:
- :
- Power Supply: AC220V±10% 50/60Hz; 110V±10% 60Hz; 400VA
- :
- : Package Size
- 670*590*790mm(W*D*H):
- :
- Gross Weight: 53kg
- :
- : Reagent Package Size
- 320*320*290mm (W*D*H):
- :
- Reagent Gross Weight: 22kg
- :
- :